Sorry, you need to enable JavaScript to visit this website.
hamburger overlay

Question
Why must the visual color check be realized within 30 to 60 seconds after application?

The Accu-Chek Active meter detects the color change in the test pad area. If the test pad is checked too early, the glucose reaction may not yet be finished. If it is done too late, the color may have already changed due to drying.