No. As seen with almost all products designed for at-home use, the Accu‑Chek Instant S meters do not have a claim for diagnosis.
During our human factors testing, we discovered that there could be some confusion in misinterpreting an error with an actual test result if the meter is held upside down while viewing the display. We implemented this approach to showing error codes so as to eliminate any risk of misinterpretation.
At least 720 readings can be stored inside the meter. However only the last result and the 7, 30 and 90-day averages can be reviewed on the meter. To review all values the meter must be connected to one of the software solutions (e.g. Accu‑Chek Smart Pix or Accu‑Chek Connect).
Performing a control test lets the user know the meter and test strips are working properly. A control test should be performed when: • A new test strip box is opened • The test strip container has been left open • The test strips are thought to be damaged • Checking the meter and test strips • The test strips were stored in extreme temperatures, humidity, or both • The meter was dropped • The test result does not match how the PwD feels • Checking to see if the PwD is performing the test correctly
The Accu‑Chek Instant S meters have a fast test time of less than four seconds.
Test results may be seen in 7, 30 and 90-day averages.
Flagging of test results is not possible on the meter itself. However, if a patient chooses to do so, they can add this information via Accu‑Chek Connect Online account.
No, the USB port for any of the Accu‑Chek Instant S is not water- or dust-resistant.
No, the USB cable does not charge the meter’s batteries. Remove the USB cable after downloading results. In fact, leaving the USB cable connected to the meter and PC after download could draw power from the meter batteries. A blood glucose test cannot be performed while the Accu‑Chek Instant S meter is connected to a PC with a USB cable.
The meters will retain the date and time settings for up to 2 hours.
If replaced within a reasonable time, the memory is not lost when the battery is replaced. If date and time no longer appear on the meter’s display after a battery change, a sync with the DMS is required.
With the Accu‑Chek Instant S meter, it is not necessary to delete stored results. Once 720 blood glucose results are in the memory; adding a new result causes the oldest one to be deleted. Additional editing solutions for results are available with the Accu‑Chek Connect Online system and the Accu‑Chek Smart Pix system.
Accu‑Chek Connect Online or Accu‑Chek Smart Pix: For Instant S meter, the date and time will be set and displayed during the process of setting up an Accu‑Chek Connect Online account via USB cable. The date and time will match that of the connected device. On subsequent data transfers, the date and time will automatically sync to the connected device date and time if there is a difference of more than 10 minutes. The date and time will also be set when configuring the meter with the Accu‑Chek Smart Pix system